SAKAGUCHI RDXL6SD Thermocouple x4/RTD x2, temperature data logger
SAKAGUCHI
RDXL6SD
Thermocouple x4/RTD x2, temperature data logger
Specifications:
| model number | RDXL6SD | |
| Input | 4-channel input (one of the following) Miniature thermocouple connector, 3-pole Pt100 (3.5 mm connector) | |
| Input temperature range | J thermocouple | -200 to 1200°C | 
| K thermocouple | -200 to 1372°C | |
| T thermocouple | -200 to 400°C | |
| R thermocouple | 0〜1768°C | |
| S thermocouple | 0〜1768°C | |
| N thermocouple | 0 to 1300°C | |
| E thermocouple | -200 to 1000°C | |
| RTD | -200~850°C (you need to purchase RTD connector model RDXL4SD-CONN) | |
| Accuracy (Temperature) | Thermocouple: | ±0.1% or 0.8°C | 
| Resistance thermometer | ±1.0% or 1.0°C | |
| Resolution (temperature) | 0.1° below 1000℃, 1° above 1000℃ | |
| Display | TFT LCD (touch panel, with backlight) 72 mm, 320 x 240 pixels | |
| Configurable parameters | Temperature unit, alarm, signal processing, date and time, data logging, power options, graph channel | |
| Alarm setting | 12 alarms (2 channels) Can be set individually with HI/LO | |
| Signal processing | Average value, minimum value, maximum value, standard deviation, temperature difference of 2 channels | |
| Response time | 1 second | |
| Operating environment temperature | 0 to 50°C | |
| Power supply | 5 Vdc adapter (included) | |
| Current retraction | 250mA in full display mode, 120mA in energy-saving mode | |
| weight | 180 g (not including batteries) | |
| Dimensions | 136 W x 71 H x 32 D mm | |
| Pt100 Connector (RTD) | 3-pole, 3.5 mm connector | |
 
                     
                                                                         
                     